Article

The CONTRACT Study: Navvus Fractional Flow Reserve to Reduce Contrast, Cost and Radiation

Permissions
Permissions× For commercial reprint enquiries please contact Springer Healthcare: ReprintsWarehouse@springernature.com.

For permissions and non-commercial reprint enquiries, please visit Copyright.com to start a request.

For author reprints, please email rob.barclay@radcliffe-group.com.
Information image
Average (ratings)
No ratings
Your rating

Abstract

A satellite symposium, sponsored by ACIST® Medical Systems, was held at EuroPCR 2016 to introduce new technologies in fractional flow reserve (FFR) and intravascular ultrasound (IVUS). FFR is the mainstay of functional haemodynamic assessment of coronary artery lesions, guiding decisions in percutaneous interventions (PCIs). The new RXi™ Rapid Exchange FFR system, featuring an ultra-thin monorail pressure microcatheter (Navvus®) has the potential to simplify PCI procedures as well as to allow the use of a workhorse coronary wire to facilitate both complex and routine FFR assessments. The system has been shown to be feasible and safe in everyday clinical practice and clinical studies. This session also introduced the first 60 MHz high definition (HDi™) IVUS system, which provides improved image quality over traditional IVUS systems, and is able to visualise bioresorbable vascular scaffold (BVS) without the need for contrast clearing.

Received:

Accepted:

Published online:

Support:The publication of this article was supported by ACIST Medical Systems.

Copyright Statement:

The copyright in this work belongs to Radcliffe Medical Media. Only articles clearly marked with the CC BY-NC logo are published with the Creative Commons by Attribution Licence. The CC BY-NC option was not available for Radcliffe journals before 1 January 2019. Articles marked ‘Open Access’ but not marked ‘CC BY-NC’ are made freely accessible at the time of publication but are subject to standard copyright law regarding reproduction and distribution. Permission is required for reuse of this content.

Dr Daemen began by describing the new ACIST RXi Rapid Exchange FFR system, which utilises an ultra-thin monorail pressure microcatheter (Navvus, ACIST Medical Systems) and is illustrated in Figure 1. This unique microcatheter allows the use of a guidewire of choice, and may facilitate negotiating complex anatomies, as well as allowing multiple pullback measurements while maintaining guidewire position.

The Navvus FFR to Reduce Contrast, Cost and Radiation (CONTRACT) study was an investigator-initiated, single-centre prospective, observational cohort study that aimed to assess whether the use of the RXi Rapid Exchange FFR device reduces cost, contrast and radiation compared with conventional wire-based FFR systems. The study enrolled all consecutive patients referred to the centre for coronary angiography or PCI and an indication to perform FFR at the discretion of the operator, between 1 September 2014 and 28 February 2015. Patients undergoing a procedure on odd dates (i.e. 1st, 3rd, 5th, etc.) were assigned to a conventional FFR system, while those on even dates (2nd, 4th, 6th, etc.) were assigned to the RXi Rapid Exchange FFR system (Navvus group). Of the 238 patients enrolled, 97 patients underwent FFR using the Navvus microcatheter. Of these, 15 patients required the use of two or more devices: in three cases the microcatheter did not cross the lesion, a device defect occurred in eight cases (a manufacturing issue that has since been resolved), the result was questioned in one and unclear in two. The remainder of the patients (n=141) were assigned to conventional FFR (cFFR): St Jude Medical Aeris/Certus (n=120), Radi Medical (n=11) and Volcano Prime/ComboWire® (n=10). Of these, five patients required the use of two or more devices: one was due to a connection problem, two were due to a device defect, one because of unsterile equipment and one because the results were unclear. Final analyses were performed on the cohort in which only one device as used (82 in the Navvus group and 136 in the cFFR group).

There were no significant differences in baseline risk profile and procedural characteristics with the exception of a higher number of small vessels in the Navvus group. Multivessel disease was present in 52.4 % in the Navvus group versus 40.4 % in the cFFR group (p=0.09), leading to multivessel FFR in 22.0 versus 24.2 % (p=0.70) in both groups, respectively. Not including the cFFR wire, the mean number of additional guidewires used was higher in the Navvus group as compared with the cFFR group (2.5 ± 1.2 versus 2.0 ± 1.8, respectively; p=0.03). However, when disregarding the first wire needed to use the Navvus system, the average number of additional wires needed was lower when the Navvus system was used, namely 1.5 versus 2.0; p=0.02.

The mean number of FFR measurements was comparable between both cohorts (1.57 ± 0.84 versus 1.52 ± 0.77; p=0.69) in the Navvus and cFFR group, respectively. Mean FFR values were slightly, though non-significantly, lower in the Navvus as compared to the cFFR group (0.82 ± 0.08 versus 0.83 ± 0.10, respectively; p=0.42), leading to a somewhat higher number of positive FFR measurements in the Navvus group (51.9 %) as compared to the cFFR group (43.8 %; p=0.25).

Catheter-based Relative Risk Reduction: The Navvus Microcatheter

Article image

The CONTRACT Study: Procedural Times of the Navvus System Compared with Conventional Fractional Flow Reserve

Article image

The procedural costs were €1,994 in the Navvus group, compared with €1,930 in the cFFR group (p=non-significant). Subgroup analysis revealed that the use of the Navvus system was associated with numerically (though not significantly) lower costs in patients with positive FFR measurement and those with multivessel disease. The use of contrast agent was similar across the patient population, but a non-significant lower usage of contrast agent was reported in the Navvus group in patients undergoing multivessel FFR and with multivessel disease. Use of radiation was slightly higher in the Navvus group, but again the difference was non-significant. Finally, the procedural time was slightly lower in the Navvus group but once more differences were non-significant (see Figure 2). In a multivariate analysis, significant predictors of cost included bifurcation (p=0.001), number of stents, number of balloons and use of optical coherence tomography (OCT) (all p<0.001), but when correcting for these, use of the Navvus system did not result in a reduction of cost. The same was found with contrast and radiation use.

The investigators concluded that the RXi Navvus FFR system, as compared to conventional pressure wire-based FFR in daily clinical practice, was associated with comparable procedural costs, amounts of radiation and contrast used. A trend was seen to potential advantages in patients with multivessel disease or positive FFR, in which the use of the Navvus system was associated with lower costs, contrast and a shorter procedure time. The study had several limitations: it was a single-centre, non-randomised study with a limited number of patients. In addition, the final analyses were only performed in procedures in which only one device was used. The cost analyses based on local list prices and the potential specific technical advantages of the system were not assessed. Nevertheless, the results indicated that the use of the RXi Navvus FFR system is feasible and safe in everyday clinical practice.

The FFR-Stent Evaluated at Rotterdam Cardiology Hospital (SEARCH) Registry has recently commenced recruiting. This registry aims to determine the impact of FFR values post-PCI on long-term clinical outcomes in an all-comer PCI setting. An FFR assessment using the ACIST Navvus Rapid Exchange FFR microcatheter system will be performed after each PCI. The FFR value distal to the most distal stent and a pullback evaluation will be performed. The primary outcome measure is a composite of major adverse coronary events, including all-cause death, myocardial infarction (MI) and target vessel revascularisation (TVR) at 30-days and yearly up to 5 years. To date, 240 of an intended 1,000 patients have been recruited.

References

  1. Levine GN, Bates ER, Blankenship JC, et al. 2011 ACCF/AHA/ SCAI Guideline for Percutaneous Coronary Intervention: a report of the American College of Cardiology Foundation/ American Heart Association Task Force on Practice Guidelines and the Society for Cardiovascular Angiography and Interventions. Circulation 2011;124:2574–609.
    Crossref | PubMed
  2. Task Force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS), European Association for Percutaneous Cardiovascular Cardiovascular Interventions (EAPCI), Wijns W, Kolh P, Danchin N, et al. Guidelines on myocardial revascularization. Eur Heart J 2010;31:2501–55.
    Crossref | PubMed
  3. Diletti R, Van Mieghem NM, Valgimigli M, et al. Rapid exchange ultra-thin microcatheter using fibre-optic sensing technology for measurement of intracoronary fractional flow reserve. EuroIntervention 2015;11:428–32.
    Crossref | PubMed
  4. Jang JS, Song YJ, Kang W, et al. Intravascular ultrasoundguided implantation of drug-eluting stents to improve outcome: a meta-analysis. JACC Cardiovasc Interv 2014;7:233–43.
    Crossref | PubMed
  5. Fearon WF, Fractional Flow Reserve-Guided Percutaneous Coronary Intervention, In: Price MJ (ed). Coronary Stenting: A Companion to Topol's Textbook of Interventional Cardiology. London: Elsevier, 2013.
  6. van Nunen LX, Zimmermann FM, Tonino PA, et al. Fractional flow reserve versus angiography for guidance of PCI in patients with multivessel coronary artery disease (FAME): 5-year follow-up of a randomised controlled trial. Lancet 2015;386:1853–60.
    Crossref | PubMed
  7. Nam CW, Mangiacapra F, Entjes R, et al. Functional SYNTAX score for risk assessment in multivessel coronary artery disease. J Am Coll Cardiol 2011;58:1211–8.
    Crossref | PubMed
  8. Menon M, Jaffe W, Watson T, Webster M. Assessment of coronary fractional flow reserve using a monorail pressure catheter: the first-in-human ACCESS-NZ trial. EuroIntervention 2015;11:257–63.
    Crossref | PubMed
  9. NCT02577484, Assessment of Catheter-based Interrogation and Standard Techniques for Fractional Flow Reserve Measurement (ACIST-FFR). Available at: https://www.clinicaltrials.gov/ct2/show/NCT02577484 (accessed 21 May 2016).
  10. Johnson NP, Jeremias A, Zimmermann FM, et al. Continuum of vasodilator stress from rest to contrast medium to adenosine hyperemia for fractional flow reserve assessment. JACC Cardiovasc Interv 2016;9:757–67.
    Crossref | PubMed